Focus unit; PCON

For technicians and partly for sales managers!

 

 

 

These instructions describe the procedures to install and adjust the focus unit of the scanner Pannoramic Confocal. To help resolve problems with the focus unit a hardware description of implemented components and adjustment procedures are added.

 

·     Software relevant settings are based on the software version 1.19.

 

 

Contents

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

General

 

 

 

 

 

The described focus unit is a component of the scanner Pannoramic Confocal and includes the following main parts:

 

 

The focus unit is derived from the P250 and contains the following modifications:

 

·     Objective changer is separated from the focus unit; it is mounted separately, keeps a separate unit and description

·     To clean the condenser unit, it is constructed removable, contains an acryl glass protector and is fixed with a latch.

·     Adjustment of the optical axis is modified and might require checks after adjustments.

·     Shutter mechanics is separated and situated on the scanner plate; so the shutter arms are not required. In analogy to the other scanners, it will be discussed here.

·     Because the shutter arms are removed, the focus unit does not use hardware limits! To see and check the ex-center position, only a visualization pin is used.

 

 

Remark

Might be, there are inherit images and texts from the P250. To make all the helpful images needs time! Please be insightfully, I’m working on it!

 

 

 

 

 

 The focus unit gives the possibility of focusing the FOV (field of view, seen by the scan camera) automatically during the scan process of the sample.

 

The objective changer unit is separately mounted (not onto the focus unit) and allows the consecutive use of two, preinstalled objectives. The selection or exchange of the objective is initiated by software before a slide scan session is started. The movement of the objective holder disc is executed via a DC-motor and the position is sensed via Hall sensors. Each objective position has a separate hall sensor, so the software knows always which objective is actually in use. To guarantees the proper position of the objective in the light path, the final objective position is fixed via a form-fit mechanism.

 

The unit was developed for the use of objectives of the following types:

 

Plan-APOCHROMAT 20x/0.8      and    C-APOCHROMAT 40x/1.2W

 

 

The objective type “Apochromat” contains several, special chromatic and spherical corrections and delivers so an image of very high color trueness with very small spherical aberration.

 

If the mechanical dimensions do not exceed the size of the Plan-APOCHROMAT 40x0.95 type objective, the mechanical mounting is identical and the focus distance of the objective to the tissue is not closer then 0.25mm, other kind of objectives can be used also but it is strongly not recommended! Always check with 3DHISTECH first if a different objective should be used!

 

The shutter mechanics covers the condenser and this way the bright field illumination path is broken during fluoresce scan sessions. The two commands, “condenser cover off” and “condenser cover on” are defined by a number of steps, executed by the shutter motor.

 

Automatic scan

 

In the slide scan table of the scan program “SlideScanner.exe” you may assign the objective with which the slide should be scanned. In other words, the objective can be changed any time between two slide scanning procedures. If the slide scan table does not contain information about the objective, the scan procedure will be done with the actual objective.

 

Manual scan

 

 The objective can be changed any time, before the scan procedure is started.

 

 

 

The exchange of the focus unit is possible:

 

  • If the stepper motor or its connection is faulty
  • If the shape of any part is deformed or a part is broken.
  • If the focus unit has any fault and you are unable to fix it.

 

Requirements

  • Service program for slide scanners (SlideScannerService.exe) with actual license file
  • Slide scanner (version 1.19 or higher) and Slide scanner Viewer software (SlideScanner.exe, MViewer.exe) with actual license files or dongle
  • 1.5, 2.5, 3 and 5 mm hex key wrenches,
  • Hardware and construction knowledge of the scanner Panoramic Confocal

 

 

 

 

 

 

Attention:  Do not mix the versions of SlideScanner.exe and SlideScannerService.exe! Always use these programs with the same version number. Otherwise the SlideScannerService.exe program could produce unwanted results and SlideScanner.exe does not work correctly or even freeze!

 

 

 

 

 

 

 

 

Configure the focus unit

 

 

Since the software version 1.15 the units of the scanner are configured in the file “MicroscopeConfiguration.ini”, section [Microscope].

 

 

 

·      The path of the file MicroscopeConfiguration.ini, in the software version with the operating system Windows® 7 is:

C:\ProgramData\3DHISTECH\SlideScanner\MicroscopeConfiguration.ini

 

 

 [Microscope]

SerialNumber=PCON_xxx

MicroscopeType=3DMic10

MicroscopeSubtype=Confocal

ScanCameraType=

PreviewCameraType=CVrmc_m8_pPro

BarcodeReaderType=PreviewCamera

LoaderType=SL_1Mag_12Slide_Sensor_Horizontal2

CameraChangerType=CC_none

ReflectorTurretType=RT_None

BrightfieldLightSourceType=RGBLedLight

ObjectiveChangerType=OC_2Pos

ObjectGuideXYZType=OGXYZ_FLASH4

FlashUnitType=NoFlashUnit

NDFilterType=NDType_None

PreviewLightType=PreviewLightUnitType_Type2

ShutterMotorType=Shutter_Motor

PowerSwitchBoardType=PowerSwitchBoard_Type1

ConfocalUnitType=ConfocalUnitType_Aurox

WaterFeederType=WaterFeeder_Type1

 

 

 

 

 

 

 

 

 

 

 

 

Functional overview of the focus unit

 

 

 

Physically, the focus position is defined by the distance of the objective to the tissue. If the tissue is in the focus of the objective, a sharp image is seen by the camera. Because the tissues are different from each other in thickness, and the thickness can change inside the same tissue also, the focus position must be checked and corrected always, during the scan procedure.

 

  • In Pannoramic scanners, as well as in the PCON, the real focus position is found by moving the slide toward or away from the objective via the focus pin.

 

  • The focus position is influenced by both, the fixed (adjustable) objective position and the actual position of the focus pin.

 

Furthermore, if the slide as well as the objective will be changed, the specimen holder must be moved away from the objective to avoid collision.

 

If the shutter is changed from shutter off to shutter on or vice versa, the specimen holder is moved to a position where the focus pin is not in connection with the slide or the specimen holder. This way, the distance between specimen holder (cover slip) and objective is not affected if the ex-center crosses the upper peak of the sine wave.

During all these actions, except focusing, the objective must be protected against touching the cover slip and the focus pin and 4:1 divider must be protected against overstraining. This is realized by the focus unit, with the different positions of the focus pin and the positions of the specimen holder.

 

 

As you can see, the gaps between cover slip and objective are very small, especially if the objective C-APOCHROMAT 40x/1.2W is used.

Because both objectives can be used without adjustment of the objective position after exchange, the previously developed focus unit (the focus unit for SCAN, MIDI and DESK) must be modified for the consecutive use of the 20x and 40x objectives. The worst case is defined by the parameters of the 40x objective.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

40x objective

 

On the scale of the 40x objective, the thickness of the cover slip should be selected.

 

  • If the real thickness of the cover slip differs from the selected or adjusted value, the quality of the scanned FOV may be reduced!

 

 

 

 

 

 

 

 

 

 

 

 

 

Objective position

 

In Pannoramic type scanners the objective is mounted into the middle of the focus range, offered by the movement limits of the focus pin during focusing.

 

  • It is very important, that the objective position is adjusted well; otherwise, if the objective position does not match the focus range, offered by the focus pin movement range, a focused camera image can never be produced!
  • Because the slide thickness can vary from 0.95 to 1.05 mm, the thickness of the slide, used to adjust the objective position, is also important during the objective position adjustment.
  • See also “Adjust the objective position”. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Principle of focusing

 

The slide is hold in the specimen holder (not shown; see the X-Y-Stage). To achieve a parallel movement of the slide in relation to the objective, the specimen holding mechanics contains a parallelogram. This guarantees the position of the tissue to be always perpendicular to the optical axis during focusing. The specimen holder has a mechanical preload, so the slide has always a connection to the focus pin, except during a shutter on qr off operation.

To avoid collision of the objective with the slide during the “shutter on” command, the X-Y-stage is moved to a place (X=27000 and Y=18000 steps) where the focus pin is not in connection with the slide or the specimen holder; otherwise the 40x objective would collide with the slide or specimen holder.

This principle is used also if an objective exchange procedure is in progress with the objective changer.

 

 

The focus pin is moved by a stepper motor via an ex-center and a 4:1 divider. The resolution of the stepper motor with 6400 steps per revolution, the characteristics of the ex-center and the 4:1 divider allows a very fine resolution of the focus steps and so specimen focusing.

 

 

 

 

 

 

 

 

 

 

 

 Focusing algorithm

 

The focus position of the FOV is found, if the camera sees a sharp image. To reach this, the scan camera takes images in different positions of the entire focus range and the software finds the best focus interval. When this interval is found the real focus position is found by iteration of the interval.

 

Stepper motor

The focus motor has a resolution of 6400 steps / revolution.

 

 

Home1,2

The Home position of the entire mechanics is situated inside the focus range. Because the focus motor movement is not limited in the PCON (there are no hardware limiters) the rotor of the focus motor may rotate endless.

 

Important

 

To reach the position Home2, please remove the slide from the specimen holder and move the specimen holder into a position, where a collision with the objective is impossible to protect the objective.

 

·     The command Home2 may rotate the rotor by more 100 revolutions until the position Home2 will be found. Please take this into account, if you are working with the service program.

 

See also:      Sensors Home1,2

 

 

 

 

 

 

 

 

 

 

 

 

Visualization pin is in slot position

 

This is a special position for adjusting the physical ex-center position on the motor axle. This adjustment defines the focus range; the sine wave crosses the X-axis at 500 motor steps (see above).

 

  • Use the position of the slot in the ex-center and focus plate mounting to adjust the visualization pin into the slot position.

 

 

 

 

Remark

Instead of the two shutter arms only one visualization pin is used!

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Focus range

 

In earlier developed scanner systems (DESK, MIDI, SCAN) the step number of the stepper motor was identical with the actual focus position and this was displayed by the software. The focus range in these systems is 800 steps and the positions of the ex-center in relation to the displayed focus range is identical; the physical number of steps is the same as the displayed focus position in the software.

 

The nominal focus position is at 1600 steps

 

Because the resolution of the motor is increased by twice (1 revolution of the motor axle is done in 6400 steps), the focus range is now 1600 steps, in the PCON.

Because the focus range is physically counted from -300 steps to +1300 steps, and negative numbers as focus position are unusual, the entire focus range is shifted in the software to the focus range from 1050 to 2650; add to the physical step number 1350. 

 

In the PCON:    SW focus position = HW stepper motor position + 1350.

 

The nominal focus position is at 1850 steps

 

 

 

 

 

Nominal focus position

 

Precalculated values for possible slide thicknesses

Slide thickness [mm]

Motor position [steps]

Live view

1.05

250

1600

1.00

500

1850

0.95

750

2000

Because the specimen holder of the scanner PCON can hold slides with a thickness of 0.95mm to 1.05mm, the nominal focus position of 500 steps is defined for a slide thickness of 1.00mm. If the slide thickness used to adjust the objective position deviates from the nominal thickness, please set the focus unit motor to the relevant step number, depending on the used slide thickness. The objective position should be adjusted at the relevant focus position.

 

  • Measure the thickness of the slide with tissue to be used (but without the cover slip); use the caliper.

 

·     If the thickness of the slide deviates from 1.00mm, calculate the focus position. A deviation in thickness of 0.01mm results in a deviation of 50 motor steps.

·     Insert the measured slide into the specimen holder and set the appropriate (calculated) focus position in the live view of the program SlideScanner.exe.

·     Adjust the objective position to the preset focus position until a focused image can be seen in the live view.

 

See also:      Check or adjust the objective position

 

 

 

 

 

 

 

 

 

 

 

 

Components and mechanical construction

 

 

 

Stepper motor

 

All stepper motor relevant information, like step number and direction; go to Home1,2 or shutter on and off and the state of the sensors Home1 and Home2 are transferred via the stepper motor cables; these are connected to the “Z” connectors of the “X-Y-Z-motor” controller unit.

 

The focus motor of the PCON has a resolution of 6400 steps / revolution; the solution of the sensors Home1 and Home2 is not modified.

 

 

Because the mechanical shutter is removed from the focus unit the shutter wire does also not exist the focus unit has no hardware limits! So, the stepper motor may rotate endless. To show the momentarily position of the rotor, a so-called visualization pin is mounted on the ex-center.

 

 

See also:      Power and control”, “Step motor types” “Focus motor” and “Sensors Home1 and Home2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ex-center

 

The rotor axle of the stepper motor rotates the ex-center on which the focus plate is mounted and this moves the 4:1 divider.

 

Focus plate and 4:1 divider

 

The connection between focus plate and divider is realized with a ball-headed joining because the ex-center does an up-down and additionally a left-right movement.

 

The connection spring ensures a slippage-free connection between focus plate and 4:1 divider.

 

 

 

Attention

The mountings of the focus plate should not be loosened; here the position of the focus plate and so the position of the 4:1 divider and the position of the focus pin is adjusted!

Because the movement of the focus pin is very small, the adjustment of the focus pin position is delicate; the focus plate mounting bolts position must not be altered.

·      The adjustment of the focus pin position can be done in the factory only!

 

 

 

 

 

 

 

 

 

 

 

4:1 divider, focus pin and shutter

 

On the other end of the 4:1 divider the focus pin is realized. The focusing of the tissue is done by moving the focus pin toward or away from the objective; the mechanical construction allows a focus pin movement of 0.3 mm over all. Because the focus pin touches the slide always during focusing, the distance of the slide with tissue to the objective is modified until the focus position is found. The focus pin movement between the focus limits is limited to be 0.15 mm around the nominal focus position (the 500 motor steps position).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condenser unit

 

 

 

 

 

 

 

 

 

 

To clean the acryl glass protector, and the condenser if required, the entire condenser unit may be removed.

 

·     The acryl glass protector protects the brightfield preview illumination from the immersion liquid.

·     Because it needs to be cleaned from time to time, the condenser unit is constructed removeable.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condenser unit mounting

 

 

 

 

Remove

·     Loosen the fixing (the latch) and pull the entire condenser unit to the right

 

Mount

·     Bring the latch in nearly horizontal position (Loosen) and insert the condenser unit on the dovetail mounting until it stops.

·     Tighten the latch

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Remove the condenser

  •  Remove the condenser unit (optional).
  • To remove the condenser, loosen its fixing and drive the condenser out; away from the objective.

 

 

 

Mount the condenser

·       To mount the condenser, put it into its holder and drive the condenser in; toward the objective.

·       Mount the condenser unit.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Adjust the condenser position

 

 The adjustment of the condenser is important for the bright, uniform and optimal illumination of the FOV. This reduces so the exposure time of the camera and increases the quality of the scanned tissue.

The position of the condenser affects the following:

  • The size of the visible FOV (color shading)
  • The image contrast
  • The image resolution (the numerical aperture) and
  • The exposure time.

 

  • We assume that the focus unit and the objective changer are already adjusted; the position of the objective is correct.

 

  1. Use the program SlideScanner.exe and insert a slide with cover slip, create a live view and click inside the tissue.
  2. Use the tool “Auto focus”.
  3. If the tissue is in focus, find a “clean” FOV (without tissue or dust). Use the positioning tool  and click outside the tissue but inside the cover slip of the preview window in the program “SlideScanner.exe”.

 

  1. Loosen the condenser’s fixing bolt and put the condenser position adjustment tool over the collimator lens holder.

·     If you can not see the hole of the condenser position adjustment tool, calculate the exposure time and move the adjustment tool on the optical axis until the image of the drilling or a part of it can be seen in the live view.

 

  1. Rotate the condenser toward to the objective, so the start position for the adjustment will be found; the brightness will increase.

 

  1. Rotate the condenser backward and look at the live view. Beware of the condenser cover (shutter), don’t close it and don’t bend it. You will see, that the drilling of the condenser position adjustment tool comes into focus. While rotating the condenser you can also observe that the brightness decreases.

 

 

 

 

 

 

 

 

 

 

 

 

  1. If the center drilling of the adjustment tool is in focus, the condenser position is nearly found, a small correction of the condenser position have to be done, because the focus position of the drilling is not the real position of the collimator lens. Drive the condenser by about ¼ revolution more backward and stop the move of the condenser; tighten its fixing bolt.

 

  1. Check the correctness of the adjustment in the live view and by scanning a tissue or a part of it.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  • The unevenly illumination of the “clean FOV” will be corrected by software (the right border is a little bit darker than the left side).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Mechanical Shutter

 

 

 

General

 

The mechanical shutter is implemented as a part of the scanner plate and the shutter guide is situated between condenser and 4:1 divider.

 

During fluorescent scanning the shutter has to be closed and insures so a dark background. Other, unwanted fluorescent materials (e.g. painting, optics) can not reflect the fluorescent light or can not fluoresce and so they will not disturb the fluorescent view.

 

During bright field illumination the shutter must be fully open to ensure a bright and evenly illumination of the FOV.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Principle of shutter mechanics

 

 

The shutter has only two usable states shutter on (closed; BF illumination path broken) and shutter off (open). These positions are defined by a photo-coupler from beneath and a number of steps to go, between open and closed. In these positions the bright field illumination should be fully opened or fully closed respectively. All other possible positions of the shutter are not defined, therefore faulty positions and not allowed during the scan process.

 

 

 

 

 

 

Shutter guide

 

If the shutter motor axle rotates the shutter guide, the positions shutter on and shutter off are defined by the executed number of steps.

 

·     The shutter off position is defined so, that the BF preview illumination will not be covered.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

See also:      Shutter mechanics”, “Power and control”, “USB-controller”, “Mechanical shutter” “ Shutter motor” and “Shutter sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Adjustment procedures

 

Adjustments of the focus part and the shutter mechanics

 

 

 

 

Adjust the ex-center position

 

This adjustment must be done only if the ex-center was dismounted or the fixing bolt was loosened. With this adjustment the focus position is set to be +500steps nominal.

Because in this position the fixing bolt can not be loosened or tightened, we use the focus stepper motor position of 2100 steps for this adjustment.

 

 

Remark

Instead of the two shutter arms only one visualization pin is used; the shutter wire is removed!

 

 

1.     Loosen the ex-center fixing bolt so, that the ex-center can be moved on the motor axle.

 

2.     With the service program set the focus unit to Home1,2.

 

3.     Go forward to the focus motor position 2100 steps.

 

4.     Bring the ex-center into the horizontal position (the shutter arms shows to the left and to the right; use the slot of the ex-center mounting to find the proper position. Before you are tighten the ex-center fixing bolt please check the position of the shutter arm in relation to the shutter wire also. The shutter arm should touch the shutter wire nearly in the middle of the hook. To reach this, bring the ex-center in proper position on the motor axle and tighten the fixing bolt.

 

 

 

 

 

 

 

Attention

A small gap between motor mounting surface and ex-center mounting should exist, otherwise, if the ex-center mounting rubs on the motor mounting surface, the focus unit may lost rotor movements and so the scanned tissue may be out of focus or blurry, even if the focus rotor direction had changed.

 

5.     Go to the positions Home1,2.

 

6.     Go forward to the focus motor position +500 steps.

 

7.     In this position the shutter arms should showing exactly up and down. A deviation of approx. 10 steps from the vertical direction is allowed. If the deviation is too much, repeat the adjustment from step 1.

 

Remark

Instead of the two shutter arms only one visualization pin is used!

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dismount and mount the focus unit

 

 

 

 

Focus unit mounting

 

The mounting of the focus unit is realized with a dovetail; this is hold by a dovetail fixing clamp.

 

  • Rotate the screw driver clockwise to open (loosen) the dovetail clamp.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 Dismount the focus unit

 

  •  Make sure, that the X-Y-Stage is not in the X-home position, because in this position the focus unit can not be removed.

 

1.     With the service program go forward to the position 10,000 steps in +X direction (3 revolutions of the motor axle).

 

2.     Disconnect the cables of the focus motor (optional).

 

3.     Remove the condenser unit.

 

4.     Loosen the fixing bolt of the focus unit mounting; turn the screw driver clockwise!

 

5.     Remove the entire focus unit by pulling the unit backward.

 

 

Mount the focus unit

 

6.     Insert the focus unit into the dove tail of the scanner plate until the physical limit is reached; take care on the focus pin.

 

7.     Tighten the fixing bolt of the focus unit mounting; turn the screw driver counter clockwise!

 

8.     Connect the cables of the focus motor.

 

9.     Mount the condenser unit