Exchange the objective (DESK)

 

 

1.      Remove the upper and the Lower housing

2.      Remove the focus unit

 

3.      Remove the objective

 

1.      Insert the new / other objective; put the removed objective into the objective protector box

 

2.      Mount the focus unit

 

3.      Adjust the objective and focus position

 

4.      Check the condenser position

 

5.      Check / adjust the brightfield illumination with the LUT-Adjustment

 

6.      Check / adjust the scan area

 

7.      If the scan area was modified, execute the preview calibration

 

8.      Scan a tissue or a part of it and check the focus positions and the scan quality

9.      Mount the lower housing

 

10. Mount the upper housing

 

 

Cover

Construction; DESK

Exchange the objective

Slide Door

 

 

 

End