Exchange the objective (MIDI)

 

 

 

 

1.      Open and remove the shells; remove the rear cover

 

2.     Remove the focus unit

 

3.     Remove the objective

 

4.     Insert the new or other objective; put the removed objective into the objective protector box

5.     Mount the focus unit

 

6.     Adjust the objective and focus position

 

7.     Check the condenser position

 

8.     Check / adjust the brightfield illumination with the LUT-Adjustment

 

9.     Check / adjust the scan area

 

10. If the scan area was modified, execute the preview calibration

 

11. Scan a tissue or a part of it and check the focus positions

 

12. Mount the rear cover

 

13. Mount the cover shells

 

 

 

End