Exchange the X-carriage drive unit (MIDI)

 

 

 

 

 

1.      Open and remove the shells and remove the back wall

2.     Remove the focus unit

 

3.     Remove the X-Y-stage unit

 

4.     Read the passage about the carriage drive unit

 

5.     Remove the X-carriage drive unit

 

6.     Insert (mount) the new X-carriage drive unit

 

7.      Fit the HW-limits

 

8.     Check “Important

 

9.     If the limits are out of range, repeat from step 2

 

10. Mount the X-Y-stage unit

 

11. Mount the focus unit

 

12. Check the hardware limits in X-direction

 

13. Set the value of the parameter DeafultPositionX in the section “[SlideLoading]”.

 

14. If the limits are out of range, repeat from step 2

15. Define the HW limits for the X-carriage and update the appropriate parameter values in the file “MicroscopeConfiguration.ini” section [HardwareLimits].

 

16. Check the values of the scannable area X-min and X-max and correct them if necessary

 

17. If the values Xmin and / or Xmax of the scannable area were modified, execute the preview calibration.

 

18. Start the SlideScanner.exe program and check the preview; make a sample scan.

 

19. Mount the back wall

 

20. Mount the cover shells

 

 

 

 

End