Modifications in summer 2016
These modifications do not result in another scanner type, but to bundle
all the modifications, the terms SCAN_II, MIDI_II and DESK_II was created and
are used internally.
Modifications, named as �SMD_II� are done in all SCAN,
Scanners, delivered since summer 2016 including all the improvements and
modifications; appropriate replaced units, used in previous versions are no
longer delivered in produced scanners but may be delivered as spare part for
existing scanners.
Features
�
Since 2016 on the market
�
The slides are hold in magazines
with 25 slide positions
�
6 magazines can be hold in the input
and output stack during 1 session.
�
Designed for brightfield an
fluorescent scan sessions.
�
By using BF RGB illumination, the
brightfield scan process can be done with a monochrome scan camera.
�
By using a monochrome scan camera
for BF scan also, the camera changer and a separate FL scan camera is no longer
required.
�
The scan speed depends highly on the
used computer configuration and the speed of the camera.
�
Today, all references to the SCAN_II
referring to this version of scanner.
Allowed slide dimensions are defined as:
������������������������������������� Length:��������� 75.00 to 76.00 mm
������������������������������������� Width:����������� 25.00 to 26.00 mm
������������������������������������� Thickness:��� 00.95 to 01.20 mm
The SCAN exists in a full version (SCAN_II_FL) where the Reflector
turret unit is included and a more cost-effective version (SCAN _II_BF) without
the Reflector turret unit, so only brightfield scan sessions are possible.
Improvements in Pannoramic SCAN since
summer 2016
�
X-Y-Stage improved to reduce noise
and vibration
�
RGB BF illumination allows higher
scan speeds while higher resolution
�
Specimen holder modified, so slides
of single width and a thickness of max 1.20mm can be inserted
�
Focus unit modified to scan slides
with a thickness if 1.20mm
�
Using 5-Phase stepper motors reduces
vibration during scanning
�
Thanks the modification of BF
illumination, camera changer and a different FL scan camera is no longer
required.
More detailed information can
be found in the appropriate unit
Magazine, used in the Pannoramic SCAN_II
� The magazine type
is the same as in the SCAN and the P250.
� A gear on the rear
of the magazine allows the movement of the magazine in the feeder channel.
� The permanent
magnet is used to detect the presence of a magazine in the input stack and the
state �Output stack full� in the output stack (P250 only).
� By filling
magazines in the input stack and continuously removing finished magazines from
the output stack during the scan session is in progress, a �Continuous mode� of
the scan session can be performed.
� To reduce the
amount of different magazines, the construction of the magazine for SCAN and
P250 is identical, but the permanent magnet acts only in the P250.